  • Page 1
    Infusomat® Space P and Accessories Instructions for Use It is recommended that all pumps at your care unit are equipped with the same software version or in parallel Valid for software 687N use with Software M.
  • Page 2: Table Of Contents

    CONTENTS Infusomat® Space P Overview ………………..3 Symbols on Product ……………………5 Patient Safety ……………………6 Menu Structure / Navigation………………..10 Chapter 1 Operation ………………….13 1.1 Start of Infusion…………………….13 1.2 Entry With Different Combinations of Rate, VTBI (= Volume To Be Infused) and Time……………………….16 1.3 Bolus Application……………………17 1.4 Infusion Line Change and New Therapy Start …………..18 1.5 End of Infusion ………………………19…

  • Page 3: Infusomat® Space P Overview

    INFUSOMAT® SPACE P OVERVIEW I N F U S O M AT ® S PA C E P O V E R V I E W Arrow up and -down Press to reset single values Scroll through menus, change setting of numbers from to zero and switch back to Press to 0-9, answer Yes/No questions.

  • Page 4
    For vertical position push lever down and rotate either way until lever clicks into notch. Push lever for rotation. Caution: A maximum of three B. Braun Space pumps can be stacked together only in horizontal pump position when used with the PoleClamp SP.
  • Page 5: Symbols On Product

    SYMBOLS ON PRODUCT S Y M B O L S O N P R O D U C T…

  • Page 6: Patient Safety

    Use. Operation • The initial training of the Infusomat® Space P is to be performed by B. Braun sales personnel or other authorized persons. After each software update, the user is required to inform himself about the changes to the device and accessories in the instructions for use.

  • Page 7
    PATIENT SAFETY • Installation in medically used rooms must comply with the appropriate regulations (e.g. VDE 0100, VDE 0107 or IEC-publications). Observe national specifications and deviations. Caution: Operate the pump at least 25 cm from flammable anaesthetics to prevent explosion. •…
  • Page 8
    PATIENT SAFETY Other components • Use only pressure-proof and compatible disposable items (min. 2 bar/ 1500 mm Hg) to avoid influencing performance data — which would result in impairing patient safety. • Where several infusion lines are connected on one single vascular access, the possibility of the lines exerting a mutual influence over each other cannot be excluded.
  • Page 9
    PATIENT SAFETY • When ending PCA and starting it again the therapy data are set to default values. • Using the demand button also the patient is a permitted user. With the demand button only a PCA-bolus can be requested. This is limited to pre- defined doses by drug list and pump settings.
  • Page 10: Menu Structure / Navigation

    MENU STRUCTURE / NAVIGATION MENU STRUCTURE / NAVIGATION Cutline On/Off button Clear button Door open button OK button Start/Stop button Keypad with arrow up, -down, -left, -right button Bolus button Connection button All display screen shots are examples and may be different when related to an individual patient and individualized therapy.

  • Page 11
    MENU STRUCTURE / NAVIGATION Display Meaning All status information is available in the bottom line of the display. The desired information can be selected by using and will be displayed permanently thereafter (e. g. drug long name, current system pressure etc.). has been pressed while the pump is infusing.
  • Page 12
    MENU STRUCTURE / NAVIGATION Start Up Main Special Options Status Menu Menu Functions Menu Menu Line Dose Rate Occlusion Intermediate Dose selection Calculation Pressure volume Intermediate Prime ? Concentration Drug Library Data Lock amount Use last Change-over Bolus Intermediate Weight therapy ? from Rate…
  • Page 13: Chapter 1 Operation

    OPERATION Chapter 1 OPERATION 1.1 Start of Infusion • Ensure that the pump is properly installed. Check the equipment for completeness and damages. Do not attach the infusion bottle below the pump level. • Put the spike vertically into the infusion bottle. Fill the bottom part of the drop chamber by max.

  • Page 14
    OPERATION Chapter 1 Press down the opening lever. 1. Mind flow direction while inserting the infusion line (s. pictogramm) to prevent the danger of backflow. 2. Completely insert and then gently press the infusion line into the guidance and the sensors. 3.
  • Page 15
    OPERATION Chapter 1 is properly occluding the IV set and the door extension hook is not broken. If the door hook is found damaged or broken remove the pump from service. Caution: If a wrong line is selected the time until the pump goes into a pressure alarm may be prolonged.
  • Page 16: Entry With Different Combinations Of Rate, Vtbi (= Volume To Be Infused) And Time

    OPERATION Chapter 1 upstream sensor calibrates on the current existing underpressure in the line and air might pass the AirStop membrane. In this case please ensure that the the drip chamber is refilled and then open the door for recalibration of the upstream sensor.

  • Page 17: Bolus Application

    OPERATION Chapter 1 a) Target symbol is placed in front of VTBI: • Change of VTBI => Adjustment of time. Old and new target: VTBI • Change of time => Adjustment of rate. Old and new target: VTBI b) Target symbol is placed in front of time: •…

  • Page 18: Infusion Line Change And New Therapy Start

    OPERATION Chapter 1 1.4 Infusion Line Change and New Therapy Start Note: Always interrupt the patient connection before changing a line to avoid dosing errors. Never let the pump run unattended when changing the line. Check and clean the safety clamp regularly. •…

  • Page 19: End Of Infusion

    OPERATION Chapter 1 Remove emergency aperture cover. Turn crank inside aperture to open the door. • Close the pump door, confirm the inserted line with and open the roller clamp. • If required, prime the pump with u. Then press to proceed.

  • Page 20: Standby Mode

    SPECIAL FUNCTIONS Chapter 1 1.6 Standby Mode In the case of extended interruption, the user has the option to maintain the set values. • Press to stop the infusion. Then press for less than 3 sec. • Confirm that the pump is supposed to switch to standby by pressing u. •…

  • Page 21: Chapter 2 Advanced Operations

    SPECIAL FUNCTIONS Chapter 2 ADVANCED OPERATIONS 2.1 Status Request of Pump when Infusion is Running Press to switch between run display and Main Menu while the device is infusing. Navigate through the menu using to check parameters. In order to check the menu parameters in the Status-/Options Menu, select «Status»…

  • Page 22: Chapter 3 Special Functions

    SPECIAL FUNCTIONS Chapter 3 SPECIAL FUNCTIONS 3.1 Dosing Units and Dose Rate Calculation (Overview) The following list shows the units used in the pump: Gram family: ng, mcg, mg, g Unit family: mIU, IU, kIU, MIU Equivalents family: mEq Mole family: mmol Kilocalorie family: kcal Millitliter family:…

  • Page 23: Drug Library

    SPECIAL FUNCTIONS Chapter 3 a Enter the concentration by entering the amount of the active ingredient and the volume. In order to do so set the values with and confirm with k. a If the patient’s weight does not need to be entered press l. to choose “weight”…

  • Page 24
    SPECIAL FUNCTIONS Chapter 3 Loading a drug (including the according parameters) from the Main Menu: • Go to Special Functions Menu and press l. • Open the drug library by pressing l. • Navigate through the list with and select the care unit with l. If you have already set the care unit once on your pump this step will be skipped for the next time.
  • Page 25
    SPECIAL FUNCTIONS Chapter 3 Hard Limits: If the set rate/dose/bolus volume and bolus rate exceed the values stored in the drug library (hard limits), the drug will be rejected, a hint will be displayed and the pump will fall back into the drug selection. If this occurs while the pump is infusing the pump will continue to administrate.
  • Page 26: Patient Controlled Analgesia (Pca)

    SPECIAL FUNCTIONS Chapter 3 3.4 Patient Controlled Analgesia (PCA) For PCA a drug list with at least one drug activating the profile PCA is necessary. By this the conditions for an effective and safe therapy are defined. Switch on pump with and wait until self-check is finished.

  • Page 27: Barcoding

    SPECIAL FUNCTIONS Chapter 3 The A/D-ratio indicates the percentage of administered and demanded boli thus giving an idea about the effectivity of the therapy. An acoustic confirmation of demanded boli can be activated and modulated by in Data Lock 3. Is a demand button connected, the therapy symbol looks like this: In case there is no demand button connected the therapy symbol looks like this:…

  • Page 28: Piggyback Function

    SPECIAL FUNCTIONS Chapter 3 Piggyback Function The piggyback-mode offers the possibility to interrupt the current (primary) infusion temporarily in order to administer a piggyback (secondary) infusion. Above the pump the piggyback-infusion line is connected with a Y-connector to the administration set. The secondary infusion is supposed to be located approx. 20 cm higher than the primary infusion.

  • Page 29
    SPECIAL FUNCTIONS Chapter 3 The piggyback therapy can be repeated many times by changing the piggyback medicament or by reset of the piggyback medicament. • Go to „Set new Piggyback” in the Special Functions Menu and confirm with l. Note: The Piggyback infusion can be deleted by starting a new Piggyback and by pressing cancel and answering the question „Use last therapy“…
  • Page 30: Ramp And Taper Mode

    SPECIAL FUNCTIONS Chapter 3 3.7 Ramp and Taper Mode The Ramp and Taper Mode is designed to deliver infusions with gradual ramp up and taper down rates. The pump automatically calculates the rate increase and decrease required to match the total volume, time and ramp up/ramp down time parameters. It consists of 3 phases.

  • Page 31
    SPECIAL FUNCTIONS Starting Ramp and Taper via Drug Library: Note: Ramp and Taper settings have been configured in the Drug List Manager before and have been uploaded into the pump. • Switch on pump with and wait until self-check is finished. •…
  • Page 32
    SPECIAL FUNCTIONS Chapter 3 Taper phase The pump linearly decreases the rate in the predefined time until it reaches the KVO rate Note: After starting infusion it is only possible to change rates, time and VTBI in the continuous phase. By editing (increasing/decreasing) the plateau rate, the therapy is recalculated.
  • Page 33: Programm Mode

    SPECIAL FUNCTIONS Chapter 3 3.8 Program Mode Program Mode is for infusion requiring a non-standard delivery pattern. The user de- fines a series of intervals (max. 12 intervals) by certain parameters (rate, time, volume) for each cycle. The pump automatically gives each programmed period, one after the other Example: Program Mode should only be performed by an experienced user being familiar with the principles of the Program Mode function and properly trained in using the present…

  • Page 34
    SPECIAL FUNCTIONS Chapter 3 The settings for the steps of the infusion are shown on the display. These settings, configured in the Drug List Editor, need to be confirmed with r. • To change the values, press to change and to confirm.
  • Page 35: Intermittent Mode

    SPECIAL FUNCTIONS Chapter 3 The Main menu informs about the current interval. The configured parameters can be checked by Program Parameter Menu in Main. Intermittent Mode The Intermittent Mode consists of 2 phases. This phases will be repeated. • Bolus phase: the configured bolus is active •…

  • Page 36
    SPECIAL FUNCTIONS Chapter 3 • Select the desired drug with and press l. The pump now offers the possible therapy profiles. • Select “Intermittent Mode” with and press l. The therapy settings for “Intermittent Mode” are shown on the display. •…
  • Page 37: Dose Over Time

    SPECIAL FUNCTIONS Chapter 3 • Enter the new value and press k. The pump continues infusion. Changing the bolus after start: If the user edits the bolus the therapy progression changes. • Press • Use to select Bolus and press l. •…

  • Page 38
    Note: The feature Dose Over Time always requires the usage of dosing units (i.e., mg or mg/kg patient weight). Before using Dose Over Time contact your local B. Braun representative! Starting Dose Over Time via Drug Library: Note: Dose Over Time settings have been configured in the Drug List Manager before and have been uploaded into the pump.
  • Page 39: Chapter 4 Autoprogramming

    AUTOPROGRAMMING Chapter 4 A U TO P R O G R A M M I N G Note: All normal pump functions remain in place when orders are received via autoprogramming. The pump can accept drug orders wirelessly from the EHR system or from SpaceStation with SpaceCom.

  • Page 40
    AUTOPROGRAMMING Chapter 4 Note: Order may be cancelled prior to confirming order. • Once all values are confirmed, the Main Menu is displayed. Note: Soft Limit alert will be issued if value exceeds any soft limits set in drug library, soft limit may be overridden or value re-programmed per institutional policy.
  • Page 41
    AUTOPROGRAMMING Chapter 4 While in PIGGYback: • Message will appear on top of display indicating update is available for PRIMary. • Press key to view order. • Follow prompt, pressing to accept order or key to cancel and hold order for later.
  • Page 42
    AUTOPROGRAMMING Chapter 4 Note: Changing values on any incoming order may only be done after confir- ming all values. Once all values are confirmed you may scroll to any value and open editor with to change value. Alternately, order may be cancelled and request made for revised order to be sent.
  • Page 43: Chapter 5 Options

    OPTIONS Chapter 5 O P T I O N S The options functions may be selected and changed while the pump is infusing or stopped. To edit a menu item, select “Options” in the Main Menu and press l. Then select desired function with and follow the Instructions for Use as described.

  • Page 44: Upstream Occlusion Pressure

    OPTIONS Chapter 5 The editor is extended by maximal 3 dashes. Confirm new pressure level with and go back to Options Menu. In the Run Menu the top line shows the current infusion pressure. The bottom line and the 3 dashes before the symbol shows the pressure alarm settings.

  • Page 45
    OPTIONS Chapter 5 Level 1: A modification of values as well as a bolus application are not possible but a change of the disposable can be conducted. It is possible to navigate through all menus and status data can be checked. Starting, interrupting and switching the pump off is possible.
  • Page 46: Bolus Rate

    OPTIONS Chapter 5 In order to deactivate the function, select “Off” in the data lock, press k, enter the code and press again. 5.4 Bolus Rate • Open bolus rate in Options Menu with l. • Change bolus rate with and confirm setting with k.

  • Page 47: Macro Mode

    OPTIONS Chapter 5 5.9 Macro Mode The infusion rate appears larger on the display when the macro mode is activated and the pump is infusing. • Open macro mode in Options Menu with l. • Answer Yes/No question by pressing to activate the macro mode.

  • Page 48: Chapter 6 Alarms

    ALARMS Chapter 6 ALARMS The Infusomat® Space P is equipped with an audible and optical alarm signal. Alarm- Audible Optical signal Staff call User confirmation type signal Red LED Yellow LED Text Device flashes device alarm Press and follow Alarm and alarm the instruction on the code…

  • Page 49
    ALARMS Chapter 6 Display message Pre-alarm reason “VTBI near end“ The preselected volume is nearly infused. “Time near end“ The preselected time is almost over. “Battery nearly empty“ The battery is almost discharged. “KVO mode“ VTBI/time are reached and the pump continues the infusion at the KVO-rate.
  • Page 50
    ALARMS Chapter 6 When pushing on the battery cover listen for “click”. “Standby time expired“ The set standby time has ended. Set new time or continue with previously set therapy. “No battery inserted“ It is not possible to use the pump without a battery pack.
  • Page 51: Reminder Alarms

    ALARMS Chapter 6 «Flow» Drop chamber is completely filled or leak in the system. Examine the line for damage and check the drop chamber. “Data were reset” Therapy and pump settings could not be restored. Enter therapy again. «Therapy data were reset» Therapy data could not be restored.

  • Page 52: Alarm Hints

    ALARMS Chapter 6 6.4 Alarm Hints If inproper entries are made the display states corresponding hints (e.g. “Bol.rate out of range”; “Download failed”; “The parameter can not be modified”). These hints disappear after a few seconds and don’t need to be confirmed.

  • Page 53: Battery Operation And Maintenance

    BATTERY OPERATION AND MAINTENANCE Chapter 7 BATTERY OPERATION AND MAINTENANCE The battery has an operating lifetime of 4 hours at 100 ml/h when new. For optimal treatment of the battery, the device is equipped with protection against overcharge and deep depletion. The battery pack is charged by the pump during connection to mains.

  • Page 54
    BATTERY OPERATION AND MAINTENANCE Chapter 7 • ambient temperature • varying load (e.g. frequent boluses). The optimal life time of a battery pack will only be reached if it’s completely dischar- ged from time to time. A maintenance mode which conducts this battery maintenance is built in.
  • Page 55: Start Up Graphs And Trumpet Curves

    START UP GRAPHS AND TRUMPET CURVES Chapter 8 START UP GRAPHS AND TRUMPET CURVES Start Up Graphs Trumpet Curves The graphs show the accuracy/uniformity of flow in relation to time. They allow for the following: The delivery behaviour or delivery precision is essentially influenced by the type of the disposable used.

  • Page 56: Technical Data

    Moisture protection IP 22 (fluid protected for horizontal usage) External power supply: • Rated voltage Via B. Braun SpaceStation or optional mains adaptor (rated voltage 100 … 240 V AC~, 50/60 Hz) for stand alone operation • External low voltage 11 ……

  • Page 57
    Delivery rate < 10 ml/h: KVO-rate 1 ml/h Delivery rate < 1 ml/h: KVO-rate = set rate (default setting 0.1 ml/h) Computer connection USB connection in combination with B. Braun interface lead CAN SP (8713230) including electrical insulation. Please pay attention to safety notices. Air detector Technical sensitivity: Detection of air bubbles ≥…
  • Page 58
    TECHNICAL DATA Chapter 9 Sensitivity upstream sensor 9 levels from -120 mbar to -200 mbar (pressure reduction) Occlusion alarm pressures 9 levels up to 1.2 bar Occlusion pressure Time to occlusion alarm [min] at rate Note: At a rate [bar] [1 ml/h] [25 ml/h] [100 ml/h]…
  • Page 59
    TECHNICAL DATA Chapter 9 EMC (ELECTROMAGNETIC COMPATIBILITY)
  • Page 60
    TECHNICAL DATA Chapter 9…
  • Page 61
    TECHNICAL DATA Chapter 9…
  • Page 62
    TECHNICAL DATA Chapter 9…
  • Page 63: Chapter 10 Warranty / Training / Tsc** / Service / Disinfecting / Disposal

    • the Technical Safety Checks are carried out regularly. Warranty B. Braun provides 24 months warranty, as from the date of delivery, for every Infusomat® Space P (12 months for every Battery-Pack SP). This covers repair or replacement of parts damaged as a result of design/manufacturing errors or material defects.

  • Page 64
    WARRANTY / TRAINING / TSC* / SERVICE / DISINFECTING / DISPOSAL Chapter 10 Technical Safety Check* / Service The Technical Safety Check is recommended to be carried out every 2 years and should be documented. Servicing work must be carried out exclusively by trained personnel.
  • Page 65
    Glutaral Aldehydes Glyoxal Peroxides Hydrogen Peroxide Active chlorine Sodium Hypochlorite Acid Citric Acid Note: Do not use Hexaquart® or other alkylamine containing disinfectants. Recommended: disinfectant for wiping available from B. Braun: Meliseptol® Foam pure, Melsitt 10% and Melsept SF 10%.
  • Page 66
    Disposal The pumps as well as battery packs can be returned to B. Braun for further dis- posal. When taking care of disposing of disposables as well as infusion solutions, please consider the applicable hygiene and disposal regulations.
  • Page 67: Technical Safety Chapter 11 Instructions For Use Accessory

    Comfort additionaly includes a central alarm management and alarm LEDs. PoleClamp SP (8713130) A maximum of three B. Braun Space pumps and one SpaceControl can be stacked together when used with the PoleClamp SP. For detailed instructions on secure fixation of the PoleClamp SP please refer to «Overview Infusomat®…

  • Page 68
    INSTRUCTIONS FOR USE ACCESSORY Chapter 11 Note: A maximum of three plugs can be stacked upon each other in socket P2. Drop Sensor SP (8713175) The drop sensor provides an additional safety function and is therefore particularly recommended in connection with low delivery rates (10 ml/h). The connection of the Drop Sensor SP on the pump is located at the rear of the device, in the lower left corner.
  • Page 69
    INSTRUCTIONS FOR USE ACCESSORY Chapter 11 2.) Connect CAN/USB converter to computer outlet as described in the Instructions for Use manual. Caution: The Interface Lead CAN SP is only to be used by the service department; never use while patient is connected. Note: A maximum of three plugs can be stacked upon each other in socket P2.
  • Page 70
    INSTRUCTIONS FOR USE ACCESSORY Chapter 11 Note: Test staff call signalling before every use. The Infusomat® Space P offers three different staff call operating modes. They are displayed in the signalling scheme. Consider the staff call of the hospital when choosing an operating mode. Choose the operating mode via the service program.
  • Page 71
    INSTRUCTIONS FOR USE ACCESSORY Chapter 11 P C A — A C C E S S O R I E S • Space PCA-Kit (REF 8713554) consisting of: : — Demand button — Hook and loop tape for fixation of the demand button at the patient s arm — Line fixation connection between hook and loop tape…
  • Page 72: Ordering

    ORDERING Art. No. B. Braun Infusomat® Space P (100 — 240 V) ……..8713070 Recommended accessories for the B. Braun Infusomat® Space P: SpaceStation…………………..8713140 SpaceCover Standard………………8713147 SpaceCover Comfort………………8713145 PoleClamp SP ………………….8713130 Power Supply SP EU III ………………8713110D Power Supply SP EU III 3.0m …………….8713123D Power Supply SP GB III………………8713111D…

  • Page 73
    Codan V86 P………………….43.4291 Hint: B. Braun does not have any influence on the quality of other lines. Changes in quality of those lines may have an influence on the technical characteristics of the pump. B. Braun therefore is not liable for deviations caused by disposables from other vendors than B.
  • Page 76
    Manufactured by B. Braun Melsungen AG B. Braun Melsungen AG Sparte Hospital Care 34209 Melsungen 34209 Melsungen Germany Germany Tel +49 (0) 56 61 71-0 Tel.: +49 (0) 56 61 71-0 38911866 • Drawing No. I0687700201 Fax: +49 (0) 56 61 71-20 44 Printed on pulp bleached 100 % chlorine-free www.bbraun.com…

Дозатор шприцевой Perfusor Compact S ф.B Braun, Германия

МВР Дата: Среда, 14.Окт.2009, 14:54 | Сообщение # 1

Медтехник

На заявках

Российская Федерация



.

. В этой специализированной теме собрана вся информация о шприцевом дозаторе Perfusor Compact S компании B Braun, Германия.
. Все вопросы и пояснения, тем или иным образом, касающиеся этого аппарата, на которые в теме не найдены ответы, и ответы/советы на них необходимо размещать только в этой ТЕМЕ !!! Тем самым, вы намного облегчите другим участникам и посетителям Форума поиск ответов на их вопросы. Да, и просто в получении полезной практической информации об этом аппарате, и о проблемах связанных с ним.
. ВНИМАНИЕ ! Прежде, чем изложить в этой теме своё сообщение (пост), проследите за тем, чтобы тематика вашего сообщения соответствовала ИЗНАЧАЛЬНОЙ тематике темы (в том числе по наименованию, типу/марке, модели аппарата), и чтобы информации, подобной вашей, в ней не было!
. В противном случае –
потрудитесь найти на Форуме тему с нужной вам тематикой! И ТОЛЬКО убедившись, что на Форуме нет таковой темы, вы можете создать новую, или обратиться к модераторам за помощью!

. Не пишите сообщения попусту и от эмоциональных перевозбуждений, не плодите посты-хвосты (вместо редактирования поста для добавления в него новой информации следующая прямо за ним череда только своих новых постов с этой информацией) — это всё удлиняет тему!
В будущем другим её будет
нудно читать …

. И ещё к сведению и обязательному исполнению см. по ссылке это ТРЕБОВАНИЕ (с.20)!
А также, не увлекайтесь излишним, необдуманным цитированием, и если вы специалист, то подумайте прежде: нужна ли цитата (а особенно длинная/нудная, да ещё и хуже того — цепочка из цитат ) – поэтому излагайте свои мысли!

ЗДЕСЬ по ссылке познавательная информация.

Темы о проблемах со шприцевыми дозаторами (насос шприцевой) серии Perfusor — http://www.medteh.info/forum/59-0-1-0-10-1-[Perfusor]
.

 
Andrey0612 Дата: Понедельник, 09.Авг.2010, 21:26 | Сообщение # 41

Участник

инженер

На заявках

Российская Федерация



Ижевск

По-моему нет … По крайней мере я не встречал в нём «время наработки» (Для всего этого, как мне кажется, нужен кабель и сервисная программка).

Сообщение отредактировал Andrey0612Понедельник, 09.Авг.2010, 21:28

 
STaras Дата: Понедельник, 09.Авг.2010, 23:00 | Сообщение # 42

инженер-электроакустика, ультразвук

На заявках

Украина



Ровно

А если все-же поискать там EEPROM-ку?
В B|Braun perfusor compact (без s) стоит между двумя здоровыми микросхемами ATMEL 93C66 — наверняка наработка в нее пишется.


Чти вело свое – ибо не ты на нём едешь, но оно везёт тебя

 
warlock Дата: Вторник, 10.Авг.2010, 20:49 | Сообщение # 43

Участник

сервис

На заявках

Российская Федерация



Москва

смотрю мануал по данному девайсу, скаченный с этого сайта — BBraun_Perfusor_Compact_S
на странице 16 — расширенный тест включения
в нем пример — 0063 — operating hours — может это и есть время наработки?
тока оно четырехзначное — 20440 — не умещается

 
Andrey0612 Дата: Четверг, 12.Авг.2010, 21:50 | Сообщение # 44

Участник

инженер

На заявках

Российская Федерация



Ижевск

STaras,

Quote (STaras)

А если все-же поискать там EEPROM-ку?

В B|Braun perfusor compact s стоит ATMEL 24C64

 
STaras Дата: Четверг, 12.Авг.2010, 23:05 | Сообщение # 45

инженер-электроакустика, ультразвук

На заявках

Украина



Ровно

Quote (Andrey0612)

В B|Braun perfusor compact s стоит ATMEL 24C64

— ну и прочитать ее с рабочего и залить в ремонтируемый? Что будет?

Quote (VOL35)

сервисная программа и оригинальный кабель.

наверное как раз для этого и предназначены.
PonyProg будет достаточно.

Если самому стремно такое делать — зайди к автомагнитольщикам, они такое на раз делают.


Чти вело свое – ибо не ты на нём едешь, но оно везёт тебя

 
Andrey0612 Дата: Четверг, 12.Авг.2010, 23:35 | Сообщение # 46

Участник

инженер

На заявках

Российская Федерация



Ижевск

STaras,

Quote (STaras)

— ну и прочитать ее с рабочего и залить в ремонтируемый? Что будет?

А ты прав, идея хорошая, может стоит и попробовать.. На крайний случай, если неполучится, убедимся что выхода нет, кроме как отправлять в сервис (А нашему начальству проще вписать, чем заниматься такими вещами)

 
hateeater Дата: Понедельник, 13.Сен.2010, 20:45 | Сообщение # 47

Заглянувший

Инженер

На заявках

Российская Федерация



Новосибирск

Несколько раз попадались аппараты с такой проблемой, перезаливка дампа флешки с рабочего аппарата помогает 100%. Большую часть времени при этом занимает подпайка soic флеши к самодельному дип-переходнику wink К сожалению дампа под рукой сейчас нет, если понадобиться , поищу в архивах.

 
jazz Дата: Понедельник, 13.Сен.2010, 20:55 | Сообщение # 48

Участник

инженер

На заявках

Российская Федерация



Екатеринбург

Аккумуляторы находятся над платой клавиатуры. Со временем они текут. Плату «клинит». Снять помыть (тщательно)-и в путь.

 
Andrey0612 Дата: Понедельник, 13.Сен.2010, 22:18 | Сообщение # 49

Участник

инженер

На заявках

Российская Федерация



Ижевск

hateeater,
Если у тебя будет возможность, то посмотри.. Но можно не торопится, я пока и сам незнаю когда смогу добраться до этих аппаратов

jazz,
Тоже вариант, как получится посмотреть, так и по пробую

 
hateeater Дата: Пятница, 08.Окт.2010, 06:37 | Сообщение # 50

Заглянувший

Инженер

На заявках

Российская Федерация



Новосибирск

Добрался до дампов.
http://rghost.ru/2850225

 
Andrey0612 Дата: Понедельник, 25.Окт.2010, 15:10 | Сообщение # 51

Участник

инженер

На заявках

Российская Федерация



Ижевск

hateeater
Пишет что файл удален.. Можно ещё раз перезалить? За ранее спасибо

Сообщение отредактировал Andrey0612Понедельник, 25.Окт.2010, 21:50

 
shnyder777 Дата: Пятница, 18.Фев.2011, 16:29 | Сообщение # 52

Участник

инженер

На заявках

Российская Федерация



Брянск

После начала работы, через 2…10 секунд пищит сигнализация, на экране выводит «АААА». Движение поршня прекращается. В описании на прибор такой ошибки нет! Подскажите, кто сталкивался, на что аппарат ругается?

 
Alexium Дата: Пятница, 18.Фев.2011, 17:44 | Сообщение # 53

Инженер медтехник

На заявках

Российская Федерация



Саратов

А случем нет знака окклюзия — перечеркнутый носик шприца в верхней части дисплея? Нажатие на F покажет вторую ошибку. А так смотри головку которая толкает поршень шщприца.

 
Zagudalin Дата: Пятница, 18.Фев.2011, 18:01 | Сообщение # 54

сервисный инженер

На заявках

Российская Федерация


Question: After a type number is entered, an alarm sounds intermittently and the display alternates between «AAAA» and the most
recently used syringe.
Answer: An invalid type number was entered

 
hrit Дата: Среда, 30.Мар.2011, 21:14 | Сообщение # 55

Заглянувший

Сервис-инженер

На заявках

Российская Федерация



Владимир

Цитата medtech161 ()

118 снимается: отвести в сторону — в загрузочное положение держатель шприца, плунжер толкателя в право до упора и включить. Батарейки надо перед этим снять и 220В тоже отключить. При включении поставить.
022 — удерживая, старт включить.

046 ошибку точно также можно снять. Правда, не надолго (появляется вновь в процессе работы насоса) — по-моему это связано с окклюзией.
А что делать хрен его знает … Может кто знает да молчит?

 
medtehn Дата: Четверг, 07.Апр.2011, 20:04 | Сообщение # 56

Стажер

инженер

На заявках

Российская Федерация



кемерово

ошибка 057 при среднем положении шприца


19661116

 
Солярис Дата: Четверг, 07.Апр.2011, 20:20 | Сообщение # 57

инженер

На заявках

Российская Федерация



Набережные Челны

57
Overflow of motor step counter

 
medtehn Дата: Пятница, 08.Апр.2011, 19:00 | Сообщение # 58

Стажер

инженер

На заявках

Российская Федерация



кемерово

Получил ответ из сервисного центра:
Указанная ошибка говорит о недопустимом количестве шагов двигателя…. возможно повреждение или выход из строя как датчиков вращения так и платы привода или платы управления….

Необходим ремонт, и лучше насос отправить в сервисный центр Б.Браун или в СПб , или в Екатеринбург.


19661116

 
medtehn Дата: Пятница, 08.Апр.2011, 19:02 | Сообщение # 59

Стажер

инженер

На заявках

Российская Федерация



кемерово

кто чинил датчик вращения?


19661116

 
Alexium Дата: Пятница, 08.Апр.2011, 21:46 | Сообщение # 60

Инженер медтехник

На заявках

Российская Федерация



Саратов

Датчик вращения оптический, ничего сложного в нем нет если разбираетесь в электронике, в противном случае отдайте на ремонт «взрослым дядям».

 
kamin Дата: Пятница, 22.Апр.2011, 20:13 | Сообщение # 61

Стажер

инженер

На заявках

Российская Федерация



Уфа

В принципе,ключик мигающий не мешает работать,при нажатии на «колокольчик» перестает подавать звуковой сигнал,ключик мигает дальше и аппарат работает как обычно.Хотя ,конечно,интересно было-бы устранить это без отправки на B/Braun

 
Alexium Дата: Пятница, 22.Апр.2011, 20:30 | Сообщение # 62

Инженер медтехник

На заявках

Российская Федерация



Саратов

Вот дамп с моего BB Compact S. Советую сначала сохранить старый дамп.

 
creyt Дата: Пятница, 22.Апр.2011, 20:40 | Сообщение # 63

техник

На заявках

Российская Федерация



Кузня

Alexium, а в этом дампе какое время наработки?


у меня все диоды болят

 
Alexium Дата: Пятница, 22.Апр.2011, 20:52 | Сообщение # 64

Инженер медтехник

На заявках

Российская Федерация



Саратов

Не помню biggrin . Это принципиально?

 
creyt Дата: Пятница, 22.Апр.2011, 21:07 | Сообщение # 65

техник

На заявках

Российская Федерация



Кузня

Ну так а смысл его шить, если насос с этой прошивкой через месяц опять в сервис запросится?


у меня все диоды болят

 
Alexium Дата: Пятница, 22.Апр.2011, 21:15 | Сообщение # 66

Инженер медтехник

На заявках

Российская Федерация



Саратов

Год гарантирую.
Если лень? Сервис ББ будет рад!
Если не лень, прошейте, посмотрите наработку и сообщите всем!

 
homen Дата: Пятница, 13.Май.2011, 14:45 | Сообщение # 67

Завсегдатай

инженер

На заявках

Российская Федерация



Санкт-Петербург

Что означает «022- удерживая старт включить»? unknown

 
sever Дата: Пятница, 13.Май.2011, 17:33 | Сообщение # 68

Участник

инженер

На заявках

Российская Федерация



Красноярск

Ошибка 022 возникает если не выключив аппарат снять питание. На сколько я помню лечиться долгим нажатием на кнопку включения. 046 — это обычно проблема с контактами и шлейфом от датчика давления до платы управления. Радиотехника — это наука о контактах :)

 
Alexium Дата: Пятница, 13.Май.2011, 22:25 | Сообщение # 69

Инженер медтехник

На заявках

Российская Федерация



Саратов

046 ошибку можно попытаться сбросить разобрав головку, очистив от пластмассовой пыли и поставить ровнее белый пластмассовый рычажок.

 
homen Дата: Воскресенье, 15.Май.2011, 02:58 | Сообщение # 70

Завсегдатай

инженер

На заявках

Российская Федерация



Санкт-Петербург

Так если устранить причину ошибки, то она разве сама сбросится ?
Если нет, то расскажите подробнее как именно сбросить её, т.е. стереть из памяти ? unknown

 
Alexium Дата: Воскресенье, 15.Май.2011, 09:45 | Сообщение # 71

Инженер медтехник

На заявках

Российская Федерация



Саратов

??????.
homen я вас не понял! Эта сообщение микроконтроллера о аварийном (непредусмотренным) состоянии дозатора («считанное» с датчиков, возникшее в результате самопроверки и т.д.). Устраните это состояние и микроконтроллер о нем не будет сообщать — дозатор будет работать в нормальном режиме.
Что Вы хотели узнать???

 
homen Дата: Воскресенье, 15.Май.2011, 12:58 | Сообщение # 72

Завсегдатай

инженер

На заявках

Российская Федерация



Санкт-Петербург

я считал, что если ошибка возникла хотя бы даже на короткое время, она прописывается в eeprom, и постоянно индицируется до её сброса ?
или нет ?

 
Alexium Дата: Воскресенье, 15.Май.2011, 15:12 | Сообщение # 73

Инженер медтехник

На заявках

Российская Федерация



Саратов

В compact s нет. За Все выпущенные когда-либо по всему миру — не знаю.

 
homen Дата: Воскресенье, 15.Май.2011, 15:30 | Сообщение # 74

Завсегдатай

инженер

На заявках

Российская Федерация



Санкт-Петербург

Quote (hamster)

Добрый час! 118 снимается: отвести в сторону — в загрузочное положение держатель шприца, плунжер толкателя в право до упора и включить. Батарейки надо перед этим снять и 220 тоже отключить. При Включении поставить.
022- удерживая старт включить.

Не зря же это было написано в начале поста. Вопрос собственно именно про «022- удерживая старт включить» — что это означает ?

 
Alexium Дата: Воскресенье, 15.Май.2011, 15:51 | Сообщение # 75

Инженер медтехник

На заявках

Российская Федерация



Саратов

«Не зря же это было написано в начале поста. Вопрос собственно именно про «022- удерживая старт включить» — что это означает ?»
Я описал сброс ошибки 046 и препираться не хочу!!!
ЕСЛИ ВЫ ЗНАЕТЕ ТОЧНО ЧТО ПРИ ЭТОМ ПРОИСХОДИТ, ТО НАПИШИТЕ КАК УДАЛЯЕТСЯ ОШИБКА ИЗ EEPROM и по каким адресам она записывется, если не знаете — то я не на экзамене у ВАС.

 
SeniorJunior Дата: Вторник, 31.Май.2011, 10:41 | Сообщение # 76

ремесленник

На заявках

Латвия



Рига

Коллеги,доброго дня !

Собственно вопрос о модуле N или трансформаторе, кто-нить пробывал разбирать с целью замены предохранителя или это пустая трата времени,надо покупать у Брауна ?!

На стикере надпись, что внутри есть предохранитель на 0.1 апмпер и тепловой на 115 градусов, вот так однако :)

 
STaras Дата: Вторник, 31.Май.2011, 11:02 | Сообщение # 77

инженер-электроакустика, ультразвук

На заявках

Украина



Ровно

SeniorJunior, можно и разбирать, но проще поменять — там всего одна вторичная обмотка.

 
homen Дата: Вторник, 31.Май.2011, 11:47 | Сообщение # 78

Завсегдатай

инженер

На заявках

Российская Федерация



Санкт-Петербург

Вроде разобрать не реально, т. к. модуль полностью залит компаундом.
Можно купить и у Брауна, но наверное это не дёшево.
А можно поставить стандартный, трансформатор, главное чтобы он был герметичный, т.к. медтехника и ещё их постоянно чем нибудь поливают (по крайней мере у нас).

 
SeniorJunior Дата: Вторник, 31.Май.2011, 12:15 | Сообщение # 79

ремесленник

На заявках

Латвия



Рига

STaras, быть такого не может, как же без первичной :)

homen, вот поэтому и спросил, что залит блочек со знанимем дела :)
Если выбило предохранитель, то не факт ,что случайно, а если где-то замыкание в обмотке?!
Нет,скорее всего куплю у Брауна, мы их периодически сдаем для сброса сервисного уведомления, как-то будет не солидно выглядить самопальная конструкция :)

 
homen Дата: Вторник, 31.Май.2011, 12:24 | Сообщение # 80

Завсегдатай

инженер

На заявках

Российская Федерация



Санкт-Петербург

Согласен.
А сами сбрасывать сервисное уведомление не научились?

 
OP

Offline

Expert

Joined: May 2009

Posts: 127

Hi all,

I have a problem with bbraun perfusor compact s!On a display appears symbol «key« and on other pump appears an error code 046! How can I reset them ?Does anyone have service manual for them ?

Regards!

Recruitment / Jobs

Advertise your jobs with EBME

Joined: May 2009

Posts: 72

Scholar


Offline

Scholar

Joined: May 2009

Posts: 72

Jevremco
I’ll have to go rummaging through a lot of old docs to find what you are looking for in terms of the «key» indicator..

on the E046, the last pump I had was merely a case of fastening the syringe size potentiometer clip (Not at all in line with the error code index of the OEM)and re-calibrating the occlusion limit.

If you don’t have the correct guages I would suggest you get in touch with Braun for a set. (I found using a generic set just didn’t cut the cake «so to speak» during a previous battle with a compact s)


keeping faith is old school…..
Rather pay me first.

Joined: Apr 2005

Posts: 371

Sage


Offline

Sage

Joined: Apr 2005

Posts: 371

Checking for DAC error as indicated in the manual.
But all 3 units with same problem couldnot understand.
Thats why asking other fellow members if they have found any solutions. :-)
thanks again for your quick reply as always.
Regards
LAIQ

1.0

Error Code

1

1001 … 1013

Internal Error

2

1014

Loudspeaker not off

3

1015

Loudspeaker lost

4

1016

Loudspeaker shorted

5

1017

KuP switchoff path defect

(K_SM_CLK)

6

1018

ADC pressure out of range

7

1019

Internal Error

8

1020

FUP Flash Memory Error Software Software

9

1021

FUP different version KuP to FuP

10

1022

FUP pressure zero test fail

11

1023

FUP pressure offset test fail

12

1024

FUP EA key closed too long 20sec Keyboard defective

13

1025

Internal Error

Table 2 — 2 Device alarms of the function processor

Perfusor® Space, 1.0 gb

Definition

Loudspeaker connector

Loudspeaker

Loudspeaker connector

Loudspeaker

Loudspeaker connector

Loudspeaker

Switch off path

Pressure measurement in drive

head

Software

Pressure measurement in drive

head

Pressure measurement in drive

head

Unit Diagnosis / Calibration

Possible Cause

Check the loudspeaker connector

Check the loudspeaker

Check the loudspeaker connector

Check the loudspeaker

Check the loudspeaker connector

Check the loudspeaker

Carry out calibration

Update unit software

Update unit software

Carry out calibration

Carry out calibration

Carry out device check

2

Fault Clearance

2 — 5

Syringe Perfusor compact Braun Error 046

B|BRAUN Syringe pump Perfusor Compact | How to use | How to install and uninstall battery | FunctionПодробнее

B|BRAUN Syringe pump Perfusor Compact | How to use | How to install and uninstall battery | Function

MENGATASI ERROR 046 SYRINGE PUMP MERK B BRAUN TYPE PERFUSOR COMPACT SПодробнее

MENGATASI ERROR 046 SYRINGE PUMP MERK B BRAUN TYPE PERFUSOR COMPACT S

Насос Шприцевой BBraun Perfusor Compact SПодробнее

Насос Шприцевой BBraun Perfusor Compact S

Насос шприцевой BBraun Perfusor Compact S.Сброс ошибки 022 на шприцевом дозаторе BBraunПодробнее

Насос шприцевой BBraun Perfusor Compact S.Сброс ошибки 022 на шприцевом дозаторе BBraun

B.BRAUN Syringe pump Perfusor Compact sПодробнее

B.BRAUN Syringe pump Perfusor Compact s

perbaikan syringe pump b-braun perfusor compact sПодробнее

perbaikan syringe pump b-braun perfusor compact s

Насос Шприцевой BBraun Perfusor Compact PLUSПодробнее

Насос Шприцевой BBraun Perfusor Compact PLUS

BBraun Perfusor Compact PLUS, краткий обзор менюПодробнее

BBraun Perfusor Compact PLUS, краткий обзор меню

How to fix no power BBraun Syringe PumpПодробнее

How to fix no power BBraun Syringe Pump

Обзор инфузионного насоса B.Braun Perfusor® CompactПодробнее

Обзор инфузионного насоса B.Braun Perfusor® Compact

B|BRAUN Compact Plus Syringe Pump | How to Setup | How to use | Bolus | Stacking | Pole clampПодробнее

B|BRAUN Compact Plus Syringe Pump | How to Setup | How to use | Bolus | Stacking | Pole clamp

Медицинская аппаратура: Насос Шприцевой BBraun Perfusor Compact S. Дозы дофамина на все случаи жизниПодробнее

Медицинская аппаратура: Насос Шприцевой BBraun Perfusor Compact S. Дозы дофамина на все случаи жизни

Error 1119 Syringe Pump BBRAUN-Perfusor SpaceПодробнее

Error 1119 Syringe Pump BBRAUN-Perfusor Space

Популярное

Понравилась статья? Поделить с друзьями:
  • Перфект ворлд ошибка запуска игры
  • Песня колокольчики бубенчики звенят про ошибки моей юности
  • Песня иди один иди быстрей через ошибки
  • Песня и снова сделав ошибку я выпиваю бутылку
  • Песня жизнь любовь и ошибки всего лишь игра